EQUIPMENT OVERVIEW: Performance Specifications
2-8
DASH 2000 Patient Monitor
Revision A
2000 412-001
Respiration
Temperature (TEMP)
Invasive Blood
Pressure (BP)
Measurement technique:
Impedance variation detection
Range:
±14 Vpk /(330 pF + 20 kOhm) per elec-
trode excitation voltage/impedance
Respiration rate:
1 Ð 200 breaths per minute
Base impedance:
100 to 1000 Ohms at 52.5 kHz
excitation frequency
Detection sensitivity:
0.4 to 10 Ohms variation
Waveform display bandwidth:
0.1 to 1.8 Hz (-3 dB)
Alarms:
User-selectable upper and lower
respiration rate limits, and
user-selectable apnea limit
Number of channels:
1
Input specifications:
Probe type:
YSI Series 400
Temperature range:
0 ¡C to 45 ¡C (32 ¡F to 113 ¡F)
Resolution:
±0.1 ¡C
Output specifications:
Parameter displayed:
TP
Linearity:
<1 % from 30 ¡C to 42 ¡C
Error:
(independent of source) ±0.1 ¡C for
YSI series 400 probes
Alarms:
User-selectable upper and lower
limits for TP
Number of channels:
1
Transducer sites:
Arterial (ART), femoral artery
(FEM), pulmonary artery (PA),
central venous (CVP), right atrial
(RA), left atrial (LA), intracranial
(ICP), and special (SP)
Transducer requirements:
Excitation voltage:
±2.5 Vdc ±0.1 %
Transducer output:
50 µV/V/cmHg
Input specifications:
Range:
-25 mmHg to 300 mmHg
Offset:
±150 mmHg
Input impedance:
Common mode:
>100 k at 50/50 Hz
Differential:
>100 k from dc to 60 Hz
Output specifications:
Gain:
976 ±1 %
Frequency response:
dc to 50 Hz (+0/-3 dB)
Gain stability:
<±0.1%/¡C, and <±0.1% over any
24 hour period
Zero balance range:
±150 mmHg
Zero balance accuracy:
±1 mmHg
Zero balance drift:
±1 mmHg over 24 hours
Common mode rejection:
>60 dB at 60 Hz
Noise:
<5 mVp-p from dc to 30 Hz
Accuracy:
±2% or ±1 mmHg, whichever is
greater (exclusive of transducer)
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......