TROUBLESHOOTING: Service Tips
4-22
DASH 2000 Patient Monitor
Revision A
2000 412-001
Power Supply Fuses
The power supply has two 2-Ah fuses (slow-blow) on the primary side. If
these fuses fail, the power supply must be replaced, not the fuses. For the
secondary side, there is a 4-Ah fuse (slow-blow) on the PCB Mainboard.
If this fuse blows, the entire PCB Mainboard has to be replaced, because
the fuse is soldered onto the board.