____________________________________________________Chapter 6 – Testing and Repair, Page 6-7
Replacing UV Lamps
Overview
The UV Lamp Modules come in two lengths, a nominal 3’-0” and 5’-0”. The length and number of
modules is dependent on the length of the Ventilator section (Refer to Table T-6-7-1). There are six UV
Lamps in each Module. Lamp #1 is at the front of the UV Module and Lamp #6 is at the back of the
Module (Refer to Figure 6-7-1). If the “UVi LAMP FAILURE” yellow status light on the Ventilator is on it
indicates that one or more Lamps has failed and it must be replaced.
Table T-6-7-1
Number of UV Lamp Modules
Active
Ventilator
Length
Number
of 3’-0”
Modules
Number
of 5’-0”
Modules
4’-0” – 6’-5”
1
0
6’-6” – 7’-5”
0
1
7’-6” – 9’-11”
2
0
10’-0” – 12’-5”
1
1
12’-6” – 16’-0”
0
2
Figure 6-7-1
Typical UV Module
Front of
Module
Lamp #1
Lamp #6