5-7
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
1
Touch [Examination].
2
Touch [PW/BP Exam.].
The preset examination mode will be displayed when
the power is turned ON.
3
Enter patient information.
(Refer to "Chapter
If "Section" or "Segment" is set for [PW/BP Exam.] -
[Pulse Exam. Set.] - [Distance Input Method] in the
Settings, enter the measurement sites according to
"Entering Blood Vessel Length" (P5-20) .
4
Touch [Measurement Content].
5
Check the measurement items and
measurement sites.
1)
Check the measurement items.
Select from CAVI/Pulse Wave or ABI/Limb BP.
2)
Check the measurement sites.
Check the measurement sites.
(Refer to "Checking and Changing the
Measurement Items and Sites" (P5-19) )
3)
Touch [Back].
6
Attach the cuffs to the right brachium
and ankle.
A measurement site guide is displayed. Attach the
cuffs in the indicated position. (Refer to "Attaching the
Limb Cuffs (4-Channel Examination)" (P3-2) .)
7
Attach the PCG microphone and check
that a PCG appears in the display.
• (Refer to "Attaching the PCG Microphone" (P3-5) .)
8
Check the waveform sensitivity.
Check whether the waveform sensitivity on the home
display is appropriate or not. Usually, the sensitivity is
set to “Auto” to enable automatic selection of proper
sensitivity for each waveform. If the sensitivity is not
appropriate, you can manually select the sensitivity.
(Refer to "Setting Sensitivity Manually" (P5-20) .)
9
Press the [START] key on the display unit.
• Inflate the cuffs to acquire a pulse wave.
• If "Manual" is set as the method for acquiring
waveform, refer to "Manually Capturing and
Measuring Waveforms" (P5-24) .
• To suspend the measurement, press the [STOP] key.
If the cuff pulse wave is unstable, reattach the cuffs or
touch [Up/Down Pressure].
If the current pressure is 30mmHg or 50 mmHg, it will
be pressurized to 50mmHg or 65 mmHg, respectively.
• The pulse wave will be acquired after the cuff inflation.
• During waveform acquisition, the boundaries will be
displayed. If the boundaries are in the incorrect position,
press the [STOP] key to stop the measurement and
reattach the cuffs and PCG microphone.
• Press the [START] key to perform a real-time CAVI
check and display a CAVI indicator to check the
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...