Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
Changing the Measurement Content
Checking and Changing the Measurement Items and Sites
1
Touch [Measurement Content].
The default measurement items are the items set in [PW/BP
Exam.] - [Pulse Wave: Gen] - [Default Measurements].
2
Touch measurement items to be enabled.
The highlighted items are enabled.
3
Touch the measurement sites.
The blood pressure or pulse wave can be measured
only for the highlighted sites.
The default measurement sites are the sites set in
[PW/BP Exam.] - [Pulse Wave: Gen] - [Default
Measurement Site].
If [TBI] is not enabled as a parameter, toe blood
pressure is not measured even if [RT] and [LT] are
enabled as measurement sites, and therefore toe blood
pressure measurement results will not be displayed.
• If CAVI, ABI or TBI is selected as a measurement item,
the relevant measurement sites are selected
automatically.
• TBI is displayed only for 6-channel examination.
Setting Blood Pressure Measurements Manually
1
Set the inflation value.
The inflation value can be set manually. The sites that
can be set are the left and right ankles and left and
right toes.
Touch [Auto] to switch to manual inflation mode and
display the operation buttons.
[BP]: Enter the systolic blood pressure for the ankle.
[Inflate]: Enter the inflation value of the cuff.
*The values are set as follows:
(Blood Pressure Value) + 60 = (Inflation Value)
[Increase Pressure]:
Increases the inflation value by 20mmHg at all
sites.
[Decrease Pressure]:
Decreases the inflation value by 20mmHg at all
sites.
The inflation value of blood pressure measurement is
usually set to "Auto". Refer to P11-19 for information
on settings.
2
Touch [Back] after configuring the settings.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...