5-10
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
4
Touch [Measurement Content].
5
Check the measurement items and
measurement sites.
1)
Check the measurement items.
Select from CAVI/Pulse Wave, ABI/Limb BP or
TBI.
2)
Check the measurement sites.
3)
Touch [Back].
6
Attach the brachial, ankle and toe cuffs.
(Refer to "Attaching the Cuffs (Standard 2-Channel
Examination)" (P3-1) , "Attaching the Limb Cuffs (4-
Channel Examination)" (P3-2) and "Attaching the Toe
Cuff (6-Channel Examination)" (P3-6) )
7
Attach the PCG microphone and check
that PCG waveform appears in the
display.
• (Refer to "Attaching the PCG Microphone" (P3-5) .)
8
Check the waveform sensitivity.
Check whether the waveform sensitivity on the home
display is appropriate or not. Usually, the sensitivity is
set to “Auto” to enable automatic selection of proper
sensitivity for each waveform. If the sensitivity is not
appropriate, you can manually select the sensitivity.
(Refer to "Setting Sensitivity Manually" (P5-20) .)
9
Press the [START] key on the display
unit.
• Inflate the cuffs to acquire a pulse wave.
• If "Manual" is set as the method for acquiring a
waveform, refer to "Manually Capturing and
Measuring Waveforms" (P5-24) .
• To suspend the measurement, press the [STOP] key.
If the cuff pulse wave is unstable, reattach the cuffs or
touch [Up/down Pressure].
If the current pressure is 30mmHg or 50 mmHg, it will
be pressurized to 50mmHg or 65 mmHg respectively.
• The pulse wave will be acquired after the cuff inflation.
• During waveform acquisition, the boundaries will be
displayed. If the boundaries are in the incorrect
position, press the [STOP] key to stop the measurement
and reattach the cuffs and PCG microphone.
• Press the [START] key to perform a real-time CAVI
check and display a CAVI indicator to check the
reliability of the waveform. A red display indicates low
reliability and a green display indicates high reliability.
If the display remains red, press the [STOP] key to stop
the measurement and check whether the cuffs or PCG
microphone are attached correctly and whether the
patient is calm. (Refer to "Real Time CAVI Check"
(P5-12) .)
10
If "ON" is set for [PW/BP Exam.] - [Pulse
Exam. Set.] - [CAVI Check], the CAVI
Check display will appear after the CAVI
measurement is completed.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...