11-12
Chapter 11 Settings of This Device
Medicine - Technician
Configure settings relating to input of the medicine, subjective symptoms, comments, blood pressure measurements, body
position, respiration rate, inquiry, doctor and technician.
Medicine Info-Comment
BP Measurement-Body
*: Can be configured on the detailed settings window which will be displayed by pressing the [Details] key.
Setup Item
Description
Setting Range
<Medicine Info>
Medicine Info1 Input Info Specifies whether to enter medicine information.
ON, [OFF]
Medicine Info2 Input Info Specifies whether to enter medicine information.
ON, [OFF]
Input Length Setting
Specifies the input length of medicine codes.
2-4 [2]
Regist Code
Register a medicine code. Up to 20 codes can be registered.
(16 single-byte characters)
-
Medicine Info1
Backup*
ON: Data is backed up. The content from before the power was turned off is
called the next time the power is turned on.
OFF: Data is not backed up.
Content is erased when the power is turned off.
Daily: Recorded data is erased at the end of each day.
ON, [OFF], Daily
Medicine Info2
Backup*
ON: Data is backed up. The content from before the power was turned off is
called the next time the power is turned on.
OFF: Data is not backed up.
Content is erased when the power is turned off.
Daily: Recorded data is erased at the end of each day.
ON, [OFF], Daily
<Subjective symptoms>
Enter
Specifies whether to enter, display and print subjective symptoms.
ON, [OFF]
Input Length Setting
Specifies the input length of subjective symptom codes.
2-4 [2]
Regist Code
Register a subjective symptom code. Up to 20 codes can be registered.
(16 single-byte characters)
-
Backup*
ON: Data is backed up. The content from before the power was turned off is
called the next time the power is turned on.
OFF: Data is not backed up.
Content is erased when the power is turned off.
Daily: Recorded data is erased at the end of each day.
ON, [OFF], Daily
<Comment>
Enter
Specifies whether to enter, display and print comments.
ON, [OFF]
Input Length Setting
Specifies the input length of comment codes.
2-4 [2]
Regist Code
Register a comment code. Up to 20 codes can be registered.
(16 single-byte characters)
-
Setup Item
Description
Setting Range
Input BP
Specifies whether to enter, display and print the patient's blood pressure.
[ON], OFF
Input body position
Specifies whether to enter, display and print the patient's body position.
ON, [OFF]
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...