3-2
Chapter 3 Pre-Examination Preparations
Attaching the Cuff to the Brachium
1
Attach a brachial cuff (orange edge) to
the right or left brachium.
Wrap the cuff on the brachium with its center (where the
air hose is connected) going along the center of the inside
of the brachium and with the lower side of the cuff just
above the elbow joint.
2
Wrap the cuff around the brachium securely but not too tightly. Make sure there is no
gap between the cuff and the brachium.
Attaching the Cuff to the Ankle
1
Attach an ankle cuff (blue edge) to the
right or left ankle.
Place the cuff so that the bottom is 1cm above the ankle
and the label with the arrow is on the inside of the ankle.
Bending the knees of the patient will facilitate the cuff
application.
2
Wrap the cuff around the ankle securely but not too tightly. Make sure there is no gap
between the cuff and the ankle.
Attaching the Limb Cuffs (4-Channel Examination)
• There are right brachial, left brachial, right ankle and left ankle cuffs. Attach to the correct site according
to the edge color of the cuff.
The edge colors of the cuffs indicate the measurement site as follows.
Right Brachium: Red, Left Brachium: Yellow, Right Ankle: Black, Left Ankle: Green
• The cuff size is applicable if the INDEX (
▲
) mark is within the RANGE when attached to the patient. The
applicable circumference range is indicated on each cuff. The size range applies to the circumference
measured at the center part of the cuff.
The toe cuff can be used on either right or left side.
The cuff size and applicable size range are listed below.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...