![Fukuda VaSera VS-2000 Series Operation Manual Download Page 145](http://html.mh-extra.com/html/fukuda/vasera-vs-2000-series/vasera-vs-2000-series_operation-manual_2344837145.webp)
8-5
Chapter 8 Brachial Blood Pressure Examinations
Measuring Blood Pressure
The following is an outline of measurement operations.
For 2-channel examination, only one side (right or left brachium) can be measured. For 4-channel and 6-channel examinations,
both sides (right and left brachia) can be measured.
The measurement site can be selected from right brachium (RB), left brachium (LB) or both brachia (BB). There are two
measurement methods: manual and automatic. Manual measurements can be taken the desired number of times at the desired
intervals. Automatic measurements are taken at preset intervals.
1
Touch [Examination].
2
Touch [BP Exam.].
When the power is turned ON, the registered examination
for startup will be displayed.
3
Touch [Measurement Content] and
select the measurement site, interval
period and measurement count.
The measurement site can be selected from the right
brachium (RB), the left brachium (LB) and both brachia
(BB).
4
Attach a cuff to the right and/or left
brachium depending on the site(s) to be
measured.
• (Refer to "Attaching the Cuffs (Standard 2-Channel
5
Take the measurement.
[Manual Measurement]
Press the [START] key on the display unit to measure the
blood pressure. The blood pressure is only measured
once. The blood pressure value(s) are displayed in the
Newest area when the measurement is completed.
Press the [START] key again to take a second or third
measurement.
[Interval Measurement]
In Measurement Content, select "2 times" or "3 times" as
the measurement count and select an interval period
between 1 minute and 5 minutes. Press the "START" key
on the display unit to measure the blood pressure. The
blood pressure value(s) are displayed in the Newest area
Press
[START].
Start of
Measurement
BP Measurement
End of
Measurement
Interval
1 to 5 min.
Interval
1 to 5 min.
Print Report
Save File
Examination
Result Display
BP
Selection
1st
Measurement
2nd
Measurement
3rd
Measurement
Touch
[Use] Button.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...