![Fukuda VaSera VS-2000 Series Operation Manual Download Page 118](http://html.mh-extra.com/html/fukuda/vasera-vs-2000-series/vasera-vs-2000-series_operation-manual_2344837118.webp)
5-28
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
1) Patient information
The patient information and the examination date/time
are recorded.
2) Numeric Data Table
CAVI, ABI and TBI values, the standard range and
comments are recorded.
An ABI value of 0.90 or less is shown with a yellow
background.
3) Observation Field
Comments of the results are printed, such as an age
corresponding to the blood vessel or occlusion.
4) Blood Pressure Value
The blood pressure values and a pulse wave amplitude
graph are recorded for the brachia, ankles and toes.
5) Left Side of Standard Graph
The content to be printed can be selected by
configuring the settings.
6) Right Side of Standard Graph
The content to be printed can be selected by
configuring the settings.
7) Waveform
PCG and cuff pulse wave (for the brachia, ankles and
toes) are recorded. UT and %MAP values for the
ankles and toes are also recorded.
■
Printed Items in the Right and Left Sides of the
Standard Graph
The items to be printed can be selected from the following.
Vascular Age Graph
A graph of the standard R/L-CAVI values for each age group
is displayed. The black line indicates the average values. The
red lines indicate ±1 standard deviation. The circle marks the
position of the patient's actual age (X axis) and R/L-CAVI
value.
CAVI Trend
Previously measured CAVI values are marked on the graph.
The numbers by the marks indicate the measured values.
BP Balance
The blood pressure of each limb is indicated by the distance
from the origin. If the BP value of the right (left) brachium is
not available, the BP of left (right) brachium will be printed.
The same applies to the ankle blood pressure.
+: Indicates the median point of the systolic blood
pressure of the limbs.
↑
: The length of the arrow indicates the degree of
ischemia at the applicable site when ischemia is
suspected.
ABI Trend
A trend graph of R/L-ABI is displayed.
TBI Trend
A trend graph of R/L-TBI is displayed.
Brachial BP Trend
A trend graph of the brachial blood pressure (SYS, DIA) is
displayed.
Weight Trend
A trend graph of the patient's weight is displayed.
BMI Trend
A trend graph of the patient's BMI is displayed.
Abdominal Circumference Trend
A trend graph of the patient's abdominal circumference is
displayed.
The items to be printed can be selected in [PW/BP Exam.]
- [Print] - [Print2] -[Print Settings] in the Settings.
( ) are used if there is a possibility that the
R-CAVI and L-CAVI values may not have
been correctly measured. In this case,
check the waveforms.
( ) are used in the following cases:
1. When ABI is 0.90 or less
2. When the pulse waves of 2 beats or
less are usable (the values differ for
each beat)
(black): R-CAVI
▲
(black): Converted R-CAVI
(green): L-CAVI
▲
(green): Converted L-CAVI
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...