2-13
Chapter 2 Installation
To Turn OFF the Power
1
Check that all measurement operations
on the device and transmissions to the
computer have been completed (the
icons should no longer be blinking).
2
Remove all cables from the patient.
3
Press and hold the [POWER] key on the
display unit or the main unit for one
second.
4
When turning OFF the power of the main
unit, unplug the power cable from the
main unit. When the power is turned
OFF, the device will enter into standby
mode and the AC power indicator will
light faintly.
If the device will not be used for a long time, remove the power cable from the main unit for safety
purposes.
If the device does not turn OFF after pressing and holding the [POWER] key on the display unit, keep
pressing the [POWER] key. After pressing and holding for around 10 seconds, operations will stop
and the device will turn OFF.
The AC power indicator may stay lit for around 10 seconds after the power cable is removed. This is
not a malfunction.