22
L
N
O
Message
Cause
Countermeasure
Low BP Error
The BP value is below the measurable limit. Make sure that the cuff is properly attached. If the
error is not caused by the cuff, use other means to
perform the measurement.
Low PW Ampl.
The blood pressure could not be measured
because the pulse wave was too small.
Make sure that the cuff is properly attached. If the
error is not caused by the cuff, use other means to
perform the measurement.
Message
Cause
Countermeasure
No data selected.
A process was attempted without selecting a
data.
Select a data before performing the process.
Message
Cause
Countermeasure
Operation has timed out.
The time limit for communication with the
DMS (Data Management System) was
exceeded.
Check the connection and start the connection
again.
If this message appears frequently, contact your
local Fukuda Denshi service representative.