6-4
Chapter 6 Blood Pressure and Pulse Wave Examination: Simple Mode
7
Read the displayed precautions, then
touch [Start].
• Pulse wave acquisition starts.
• Touch [Cancel] to abort the examination. The display
will return to step 3.
Make sure to keep the patient at rest during the pulse
wave acquisition.
8
Touch [Next].
When the pulse wave acquisition is completed, the next
display appears.
If the pulse wave is obtained correctly
The device proceeds to the blood pressure examination
after 10 seconds.
The following display appears if the CAVI check is not
completed correctly.
Touch [Cancel] to perform the examination again.
To proceed to the blood pressure measurement, touch
[Next].
• The blood pressure is measured.
• When the examination is completed, the result display
appears.