21
F
I
Message
Cause
Countermeasure
Failed to connect.
Initial connection to the DMS (Data
Management System) failed.
Check the DMS status and connection to the DMS.
Also check the communication settings with the
DMS.
Failed to delete data.
Failed to delete the examination data.
There may be a problem with the external media.
Contact your local Fukuda Denshi service
representative.
Failed to format SD card.
The device failed to format the SD card.
SD card failure can be considered. Contact your
local Fukuda Denshi service representative.
Failed to format USB
mem.
The device failed to format the USB
memory.
USB memory failure can be considered. Contact
your local Fukuda Denshi service representative.
Failed to load settings.
Failed to load the settings from the SD card. Make sure that the correct SD card is used which
the settings are saved.
If the correct SD card is used, a failure of the SD
card can be considered. Contact your local Fukuda
Denshi service representative.
Failed to retrieve patient
information.
The patient information could not be
retrieved.
Make sure that the correct media is used.
Failed to save the settings. Failed to save the settings to the SD card.
There may be a problem with the SD card. Contact
your local Fukuda Denshi service representative.
Failed to save to SD card.
The device failed to save the data to the SD
card.
SD card failure can be considered. Write down the
error code at the end of the message and contact
your local Fukuda Denshi service representative.
Failed to save to Shar.
Folder.
The file could not be saved to the shared
folder.
• Check if the network connection is correct.
• Check if the shared folder settings are correct.
Failed to save to USB
mem.
The device failed to save the data to the
USB memory.
USB memory failure can be considered. Write
down the error code at the end of the message and
contact your local Fukuda Denshi service
representative.
Message
Cause
Countermeasure
Input height.
The height has not been entered.
Enter the height.
Input L as vascular length. The vascular length value L is not entered.
• Enter the vascular length value L.
• To enter this value automatically by calculating
from the patient's height, select [Height] for
"Distance Input Method" (PW/BP Examination
Settings).
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...