12-4
Chapter 12 Maintenance and Inspection
●
Performance
Inspect the PCG function using the Manufacturer maintenance - [A/D Waveform] display.
Item
Inspection Item
Procedure
Criteria
Four-Limb
Blood
Pressure
Zero point
Read the pressure value at the atmospheric
pressure when no connection is made to the
air connectors.
Should be within 0±1mmHg.
Pressure accuracy
Perform this inspection in Maintenance -
"Test Mode" (see the VS-2000 Service
Manual).
Should be within ±3 mmHg.
Maximum pressure
Perform this inspection in Maintenance -
"Pressure Test" (see the VS-2000 Service
Manual).
Should be 320±10mmHg.
Inflation time
Should be 20 sec. or less.
Air leakage
Should be 4mmHg/min or less.
Quick exhaust
Should be 10 sec. or less.
100% open
Should be within 100 to 250mmHg/sec
75% open
Should be within 20 to 190mmHg/sec
Time Limit
Should be 130±10 sec.
Should be 8±1 sec. at high speed.
PCG
PCG filter
characteristics (PWV)
Input sine-wave signals to the PCG connector
to measure the amplitude at 165 Hz and 280
Hz when the amplitude at 200 Hz is 100%.
Should be 70% or above.
PCG filter
characteristics (L)
Input sine-wave signals to the PCG connector
to measure the amplitude at 30Hz, 50Hz and
80Hz when the amplitude at 100Hz is 100%.
50% or above at 30Hz
70% or above at 50 Hz
90% or above at 80 Hz
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...