23
P
R
S
T
Message
Cause
Countermeasure
Please set DMS.
DMS is not set.
Set the DMS.
Please set SD card.
SD card is not inserted.
Insert an SD card.
Please set USB mem.
USB memory is not inserted.
Insert a USB memory.
Message
Cause
Countermeasure
Replace the backup
battery. The clock will be
initialized. (Contact your
dealer for details.)
The backup battery for the clock is running
out.
The backup battery needs to be replaced. For
details, refer to “Chapter 12 Maintenance and
Inspection”.
Message
Cause
Countermeasure
Safety Operat.
Due to abnormal increase of cuff pressure
during the measurement, or due to
measurement time-out, the safety device
activated.
Inspect the cuff and air hose for damage or
breakage.
Message
Cause
Countermeasure
The analysis range
contains some calibration
waves. Analysis cancelled.
Some calibration waves are included in the
specified analysis range.
Specify a range that does not contain calibration
waves.
The battery voltage is too
low. Cannot access Media.
During battery operation, the device could
not access the media due to low battery
level.
Connect the power cable and operate the device on
AC power, or charge the battery.
The examination data was
not saved.
The examination data was not saved when
proceeding to the next examination.
Save the examination data before proceeding to the
next examination.
There is no relevant data.
There was no data found on the DMS (Data
Management System).
Check the details of the data requested to the DMS
(Data Management System).
There is not enough space
on SD card. Please replace
the media with another
one.
There is not enough space on the SD card to
save the data.
Replace with an SD card with enough capacity.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...