11-19
Chapter 11 Settings of This Device
BP Examination
Setup Item
Description
Setting Range
BP Re-Inflation
Select whether to use the re-inflation function when measuring blood pressure. [ON], OFF
Meas Pause Right to Left
Set the pause to occur between measuring the right side and measuring the left side.
This is enabled if "ABI" is selected as the measurement item.
[5 seconds] 5-120 seconds
Measurement Pause for
Toe BP
Set the pause to occur before proceeding to the toe blood pressure measurement.
This is enabled if "TBI" is selected as the measurement item.
[5 seconds] 5-120 seconds
ABI Judgment Reference
Select the criteria to be applied to ABI measurements.
ACC/AHA Standard: Complies with the 2005 ACC/AHA Standard.
TASCII Standard: Complies with the TASCII (ACC/AHA 2011) Standard.
This is enabled if "ABI" is selected as the measurement item.
ACC/AHA Standard,
[TASCII Standard]
ABI Brachial BP Usage
Method
Specifies the brachial blood pressure usage method to be used in ABI calculations.
L/R Average: Complies with the 2000 AHA Standard.
L/R Maximum: Complies with the TASCII (2011 and 2005 ACC/AHA)
Standard.
This is enabled if "ABI" is selected as the measurement item.
L/R Average,
[L/R Maximum]
BP Measurement Method
Select whether to measure blood pressure on the left and right side separately
or in all four limbs at once.
Simultaneous Inflation: Inflates the cuffs of all four limbs at once.
Right to Left Inflation: The blood pressure on the left side is measured after
the blood pressure measurements on the right side are completed.
This is enabled if "ABI" is selected as the measurement item.
Simultaneous Inflation,
[Right to Left Inflation]
Calc auto Pressure
Increment
Set whether to automatically calculate the pressurization when measuring
blood pressure.
[ON], OFF
Toe Cuff Attachment
Guide
Select whether to enable the function for avoiding too much pressure on the
toe during toe cuff attachment (toe cuff attachment assist function).
[ON], OFF
TBI Judgment Reference
Select the TBI criteria for outputting an interpretation in the case of possible
peripheral arterial disease.
An interpretation is output when TBI<0.61 or when TBI<0.70.
This is enabled if "TBI" is selected as the measurement item.
0.61, [0.70]
Inflate Brachial Cuff for
TBI
Specifies whether to measure brachial blood pressure when measuring TBI.
ON: Brachial blood pressure is measured when measuring TBI.
OFF: Brachial blood pressure is not measured when measuring TBI. The
brachial blood pressure from ABI is used.
This is enabled if "TBI" is selected as the measurement item.
ON, [OFF]
Default Ankle Pressure
Specifies the default pressurization to be used when measuring ankle blood pressure.
[200mmHg], 100mmHg-295mmHg
Default Toe Pressure
Specifies the default pressurization to be used when measuring toe blood pressure. [200mmHg], 100mmHg-295mmHg
T
T
T
T
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...