Trueness and precision
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
99
explained by the lacking precision of the instruments involved and
does not need to cause a dispute. In such cases, one can no longer
speak of different measurement results. However, if one wishes to
prove that the difference is systematic, then the precision of the
involved instruments must be increased (e.g., by increasing the
measurement time or by decreasing the measurement distance or
by increasing the number of measurements) and in this manner the
confidence interval made narrower. If this establishes a systematic
deviation, it must be researched which of the two results is correct
and which one is not or even if both are false. How to do this will be
shown in the next section.
14.6 Trueness and precision
Trueness
Trueness is a property of measurement readings, namely to
possess larger or smaller systematic deviations from the correct
values.
As a rule, a systematic deviation is discovered when an instrument
has changed. It is a result of inadequate measurement device
monitoring if false readings are measured over longer periods.
However, the calibration of the instrument may also have been
wrong from the outset (e.g., through erroneous entries of nominal
values during the calibration or when using a wrongly labeled
calibration standard). Or the calibration standard in its composition
does not fit the properties of the instrument setup. A wrong
calibration can be corrected by re-measuring with one or more
standards and possibly through normalization or re-calibration.
Errors caused by faulty calibration standards or software influences
would become apparent only through comparative measurements
between different calibration standard sets. Both types of
deviations are systematic. Systematic measurement deviations
lead to situations that different users (e.g., supplier and customer)
measure different results at the same specimen.
Precision
In contrast to trueness, precision is an instrument property. It can
be easily characterized, e.g., by repeated measurements on a
sample (whose true value must not even be known) at one and the
same location under repeatability conditions. The mean value
produces the requested measurement reading and the standard
deviation of the test series is a measure for the instrument’s
precision. If the measurement application has been calibrated
correctly, then the mean value of the test series is also correct and
corresponds to the true value (which is the mean value that results
from infinitely repeated measurements) within the confidence
interval.
However, if one knows the correct measurement result of a sample
at a marked location and measures it with a different instrument and
by a different user, the measurement result is typically different.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...