FISCHERSCOPE
®
X-RAY 5000 Series
159
General Specification
Intended use
Energy dispersive x-ray fluorescence measuring system (EDXRF) to analyze and
measure coatings and layer systems continuously in production processes.
Element range
Sodium Na (11) to Uranium U (92) if measured in vacuum
Chlorine Cl (17) to Uranium U (92) if measured in the atmosphere
Features
Calibration
Quick and easy with a master standard directly in the production process
Remote control
Via programmed tasks or via various data interfaces
Data Interface
RS232, Command files, OLE automation, TCP/IP, PROFIBUS and PROFINET via
OPC mainly with SPS type SIMATIC, others on request
X-Ray Source
X-ray tube
Microfocus tungsten tube with berylium window, other tubes on request
High voltage
Three steps: 10 kV, 40 kV, 50 kV
Aperture (Collimator)
Fixed, adapted to the particular application
X-Ray Detection
X-ray detector
Peltier-cooled silicon-drift-detector (SDD)
Resolution
(fwhm for Mn-K
α
)
≤
140 eV
Signal processor
Digital pulse processor
Distance compensation
± 3 mm (118 mils) for measurements on glass substrates, with specific applications on
glass substrates up to ± 5 mm (197 mils)
Measuring distance
60 – 100 mm (2.4 – 3.9 in), other distances on request
Electrical data
Power supply
AC 115 V or AC 230 V 50 / 60 Hz
Power consumption
max. 100 W (without evaluation PC)
Protection class
IP50
Environmental Conditions
5100
5400
Measuring environment
Atmosphere
Vacuum
Maximum sample surface
Temperature
≤
40 °C (104 °F)
≤
400 °C (752 °F)
Cooling
–
Water cooling, e. g. with
water temperature 20 °C (68 °F),
water pressure 5 – 9 bar, flow rate 6 l/min
Operating temperature
0 °C – 40 °C (32 °F – 104 °F) around the housing
Storage temperature
0 °C – 50 °C (32 °F – 122 °F)
Admissible air humidity
≤
95%, non- condensing
Dimensions
5100
5400
Weight
37 kg (82 lb)
47 kg (104 lb)
Flange dimensions
ISO 250F
Outside W x D x H
236 x 350 x 370 mm (9.3 x 13.8 x 14.6 in)
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...