Pre-run for determining the control limits
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
109
Operation.:5 x 10 measurements
Final evaluation fromGroup No. 1 measured 12:22:09
toGroup No. 10 measured 12:39:55
nbl= 1 Ni.= 7.54
μ
m Cu.= 7.79
μ
m
s = 0.085
μ
m s = 0.098
μ
m
nbl= 2 Ni.= 7.45
μ
m Cu.= 7.70
μ
m
s = 0.053
μ
m s = 0.155
μ
m
nbl= 3 Ni.= 7.36
μ
m Cu.= 7.68
μ
m
s = 0.123
μ
m s = 0.097
μ
m
nbl= 4 Ni.= 7.43
μ
m Cu.= 7.57
μ
m
s = 0.078
μ
m s = 0.174
μ
m
nbl= 5 Ni.= 7.45
μ
m Cu.= 7.54
μ
m
s = 0.155
μ
m s = 0.093
μ
m
nbl= 6 Ni.= 7.39
μ
m Cu.= 7.61
μ
m
s = 0.171
μ
m s = 0.148
μ
m
nbl= 7 Ni.= 7.46
μ
m Cu.= 7.69
μ
m
s = 0.111
μ
m s = 0.160
μ
m
nbl= 8 Ni.= 7.35
μ
m Cu.= 7.66
μ
m
s = 0.103
μ
m s = 0.135
μ
m
nbl= 9 Ni.= 7.49
μ
m Cu.= 7.72
μ
m
s = 0.215
μ
m s = 0.168
μ
m
nbl= 10 Ni.= 7.56
μ
m Cu.= 7.84
μ
m
s = 0.158
μ
m s = 0.101
μ
m
Mean value x..
7.449
μ
m 7.681
μ
m
Standard deviation s
I
0.125
μ
m 0.133
μ
m
Confidence mean value
0.051
μ
m 0.066
μ
m
Range R(x.)
0.321
μ
m 0.327
μ
m
Lower control limit (x)
7.26
μ
m 7.46
μ
m
Upper control limit (x)
7.64
μ
m 7.90
μ
m
Upper control limit (s)
0.300
μ
m 0.338
μ
m
Number of groups
10 10
Measurement time
20 s
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...