8
FISCHERSCOPE
®
X-RAY
Safety Information
X-Radiation Safety
1.4
X-Radiation Safety
The design as a fully protected instrument and the functions of several integrated protec-
tive devices protects the operating personnel and the surroundings fully from x-radiation.
Two devices that work independent of each other ensure that x-radiation can be generated
only when the measurement chamber is closed. For the safe operation of the instrument
please also note the intended use; ref. Chap. 1.1. on Page 7.
Observe the laws and regulations of the country where the FISCHERSCOPE X-RAY
is operated. The responsible government authority must possibly be informed or
consulted about the commissioning of the FISCHERSCOPE X-RAY.
In addition, we would like to point out the following:
Operate the instrument according to the instructions in this manual and the online help.
Do not attempt any structural alterations on the instrument.
Maintenance and service work on the instrument may be performed only by FISCHER
or by companies authorized by FISCHER for this purpose.
1.4.1 Repeat Inspection
FISCHER recommends to check the instrument by an expert in intervals of no longer than
five years. Observe the laws and regulations of the country where the instrument is oper-
ated.
If you prefer to perform the repeat inspection yourself, contact the responsible authority
for the expert responsible for your area.
During the repeat inspection, the expert must be provided with the Certificate of Approval.
Store the Certificate of Approval in a manner that it will be available at the time of the
inspection.
The responsible authority must be notified of the result of the repeat inspection without
being requested to do so.
1.4.2 Safety Devices
The instrument is equipped with a connector where the safety devices must be plugged.
Safety devices are absolutely necessary to protect the operating personnel and the
surroundings fully from x-radiation. For example, a security device with sensors can be
mounted at the door of the vacuum chamber. The sensors are connected to the
X-RAY 5000. Thus, the x-radiation is switched off, when the door is opened.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...