Measurement device monitoring for the Fischerscope X-RAY
Pre-run for determining the control limits
110
FISCHERSCOPE
®
X-RAY
All printed values are available in the print form of WinFTM
PDM as
variables (including the control charts). They can re-calculate the
printed control limits by using the original values (mean value x. and
standard deviations s) and the equations specified in the previous
section. Please note that the pre-run has been concluded within a
short period (here: 17 minutes) in order to ensure the stability of the
instrument during the pre-run.
The protocol shows the video image of the reference sample that
the camera of the Fischerscope XRAY has recorded. The
measurement area is identified by an orange-colored ring, the
measurement spots where the instrument made individual
measurements during the pre-run are identified by yellow crosses.
Even though the reference area is small compared to the
dimensions of the sample, the distribution of the coatings within the
reference area can be non-uniform. Because during the check
measurement any measurement spot may be selected within the
reference area, a potential inhomogeneity of the coating
thicknesses must already be taken into account during the pre-run.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...