30
FISCHERSCOPE
®
X-RAY
User Interface of the WinFTM Software
Status Bar
5.5
Status Bar
The status bar is located at the bottom of the WinFTM window and displays the following:
5.6
mq Value
The mq value (measurement quality value) is a quality factor of the measurement value.
It is displayed in the status bar.
The mq value is a measure for the agreement of the measured spectrum with the theo-
retical spectrum. The mq value is an indication for the confidence level and the trueness
of the measurement value. The smaller the mq-value of a reading is, the more correct is
the measurement result. The optimal value for mq is 0.
General recommendation:
Display
Explanation
Help text
If you hover with the mouse pointer over a command button, addi-
tional help text will be displayed here.
Measurement
distance
Measurement distance between the specimen measurement spot
and collimator minus 15 mm, when the video image is focused.
XYZ coordi-
nates
1
1) only available at instruments equipped with motor-driven measuring stage
Coordinates of the current position of the measurement stage.
mq value
Quality factor of the last measurement value. The mq value is a
measure for the agreement of the measured spectrum with the theo-
retical spectrum, see Chap 5.6 on Page 30.
CAP
The caps lock key of the computer keyboard was pressed.
NUM
The number pad of the keyboard is activated.
mq = 0 – 5
The measurement is OK.
mq > 5
The measurement must be discarded.
Operator‘s name
Help text
Measurement distance
XYZ coordinates
mq value
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...