116
FISCHERSCOPE
®
X-RAY
Def.MA
Entering Compensation Spectra
Compensation Elements
In the area
Compensation Elements
you can enter up to four elements. The
entered compensation elements are filtered out of the spectrum. If the relevant ele-
ment area not shown in the list field, check
full Element list
.
Example tin on glass: The glass contains Strontium (Sr) in variable concentration.
If you enter Sr as compensation element, the spectral lines caused by Sr are fil-
tered out. The variable amount of Sr does not influence the measurement reading.
Compensation Spectrum
If a suitable compensation spectrum is available, you can activate the option
yes
in the in the area
Compensation Spectrum
and enter a name for the spectrum, e. g.
glass or plastics. At a later moment during the calibration you will be prompted to
position the compensation spectrum. The name entered here will then be dis-
played.
Example tin on glass: The glass causes a scattering spectrum (also referred to as
background spectrum or base spectrum). If a sample is available that is compara-
ble with the specimen and has no tin coating, you can enter here
glass
. At a later
moment during the calibration you will be prompted to position the
glass
sample.
WinFTM measures then the compensation spectrum and takes it into account
when evaluating the spectrum.
Base Spectrum
If a suitable base spectrum is available, you can activate the option
yes
in the in
the area
Compensation Spectrum
and enter a name for the spectrum, e. g. glass
or plastics. At a later moment during the calibration you will be prompted to position
the base spectrum. The name entered here will then be displayed.
Background correction
If the theoretical characteristics (measured or calculated) of the compensation
spectrum differs from the actual characteristics, you can reach a correction by
using this option. You can choose from three options: no correction, with supporting
points and by polynomial.
5. To close the window
Spectrum Background
click
OK
.
6. To close the window
Definition of measuring conditions...
click
OK
.
What you can do next
You can create a new product based on the created Def.MA, see Chap. 7.2 on Page 41.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...