Product Administration
Assigning a new measurement application to a product
44
FISCHERSCOPE
®
X-RAY
7.6
Assigning a new measurement application
to a product
A new measurement application can be assigned to an existing
product only if it has the same measurement mode as the existing
measurement function. Changes performed in the measurement
application, are effective in all products linked with this
measurement application
Procedure
1. Select in the WinFTM main menu
Product > Modify
.
The window
Modify product
appears.
2. Select the tab
Application
.
3. Click
Other Application
.
The window
Modify application
appears.
4. Select the application and click
OK
.
7.7
Modify Measurement Application
The following steps are required to change a measurement
application:
"Calibrate > Modify application" > selection of the desired
measurement application from the list of all existing measurement
applications with mouse click [OK] > changing the parameters in
accordance with the operating instruction Basic Software > [OK]:
the new parameters are transferred.
7.8
Copy measurement application
The following steps are required to copy a measurement
application:
"Calibrate > Copy application" > click on the desired application >
input of a new name or transfer of the suggestion > input of a new
number (the next new available number is suggested) > [Copy]: the
new measurement application is created.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...