Long-term monitoring
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
111
If the lower and upper limits are too far apart in reference to the
confidence interval, then the reference sample is not suitable due
to a non-uniform coating thickness within the reference area or
because the sample cannot be positioned and measured in a
reproducible fashion. It is then recommended to use a different
specimen as a reference sample.
14.11 Long-term monitoring
After the pre-run, the user is to make at regular intervals new
measurements at the reference sample, or the calibration standard,
respectively, that were used for the pre-run. For practical purposes,
the time intervals should be specified according to the difficulty of
the measurement application (see above). Each new group
provides a point for
x.
(spoken as x_bar) in the control chart and a
point for
s
(thus, the respective quality control charts are also called
x_bar / s-charts). As has already been shown above, the control
limits can also be calculated automatically by the WinFTM software
even without knowledge of the equations. One control chart each is
required for each reference sample and for each feature. This,
however, is already being taken into account by the software.
Upper and lower violations of the control limits suggest to correct
the measurement application by taking the corrective measures
recommended by us. Then a second group is measured in order to
evaluate the success of the corrective measures. If the values
continue to be outside the control limits, a new calibration may
provide help (however, with the disadvantages and limitations that
were mentioned above).
A stability test should be performed if frequent corrections become
necessary because there is suspicion that the stability of the
instrument is not ensured and maintenance or repair including
replacement of components may be necessary. In such cases, the
user should entrust his problem and help to the support and service
personnel of FISCHER.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...