Measurement Uncertainty and Measuring Ranges
Display of the measurement uncertainty
76
FISCHERSCOPE
®
X-RAY
Tab. 2: Output of the measurement uncertainty of the single
readings
Tab. 3 Output of the measurement uncertainty of the mean
values
Systematic portion
Computed from the
random error of the
calibration measure-
ment and the uncer-
tainties of the nominal
values of the stan-
dards, according to the
input in the "Calibra-
tion standard set"
Uses the error value of
the last measurement
of the block
Syst random
portion
Root of the squared
sum of the random
and the systematic
portions
Root of the squared
sum of the random
and the systematic
portions
Screen output
Only for large number display, can be switched
off in "..\Measuring head parameter\Pre-set-
tings"
Output in block print
form
Not possible
Output in final print
form
Not possible
Export
Output with every single reading, only with
Online-Export
Screen output
Not possible
Output in block print
form
ONLY FOR THE RANDOM ERROR PORTION,
regardless of the error type setting
Output in final print
form
ONLY FOR THE RANDOM ERROR PORTION,
regardless of the error type setting
Export
Output after each end of a block (block evalua-
tion), with On-line-Export. For Off-line (subse-
quently), only the random portion can be output.
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...