2.13 Calibrating and Monitoring of the Measuring Equipment
Function
Description
Calibration of new or
existing products
The obtained calibration data are stored.
Standard-free calibration
You can calibrate without calibration standards by means of
the fundamental parameter method.
Standard-based calibra-
tion with the use of up to
64 standards
You can correct the standard-free calibration by calibrating
with calibration standards. Thus, the accuracy of the anal-
ysis and thickness measurements is improved.
Matrix-specific calibration standards and pure element cali-
bration standards are available. Up to 64 calibration stan-
dards can be used for each measurement application.
Calibration standard set
Input and storing of nominal values of the used calibration
standards is performed in the
Calibration standard set
in
the measurement application of the product. After the cali-
bration has been performed, the
Calibration standard set
contains, in addition to the nominal values, the calculated
actual values including the correction. You can use this
data to create a document for monitoring the measuring
equipment.
Calculating and saving
measurement parame-
ters and calibration
results
Measurement parameters like measuring time, calibration
time, correction data, data of the last normalization and
calibration, etc. can be calculated and saved.
Library of pure element
spectra
The library of pure element spectra makes measurement of
pure elements unnecessary when performing a calibration.
This is important in cases, where pure elements are not
available.
Spectra library
During the installation of the instrument the spectra of 12 or
14 elements (depending on the detector type) are
measured and saved in the spectra library. Using these
spectra all further spectra are calculated by the instrument.
You can control whether the elements are measured or the
spectra of the spectra library are used during calibration.
Reference measurement
The Reference measurement checks and adjusts the drift
of the instrument. It is also used for monitoring the
measuring equipment.
Stability test of the tube
and of the detector
The instrument performs automatic long-term self tests to
check the primary radiation, the resolution and gain of the
spectrometer (detector and electronics).
Summary of Contents for FISCHERSCOPE X-RAY 5000 Series
Page 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Page 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Page 124: ...124 FISCHERSCOPE X RAY Calibration...
Page 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Page 183: ...WinFTM 183...