Gammapilot M FMG60 FOUNDATION Fieldbus
Accessories
Hauser
87
9.5.2
Use
Permitted
Preferred
1
2
2
3
2
2
1
2
2
2
2
1
4
4
4
4
1
1
A
B
C
D
E
A0018141
A
Level measurement, FMG60 without cooling jacket
B
Level measurement, FMG60 with cooling jacket
C
Point level measurement, FMG60 without cooling jacket
D
Point level measurement, FMG60 with cooling jacket
E
Such horizontal mounting not permitted
1
Retainer for pipe diameter 80 mm (3.15 in)
2
Retainer for water cooling jacket diameter 140 mm (5.51 in)
3
Retainer for pipe diameter 102 mm (4.72 in)
4
Bracket
For horizontal mounting with cooling jacket or collimator
(see picture E)
, the pipe
mounting must be carried out by the customer. Make sure that the clamping forces of
the mounting clamps are sufficient to prevent the FMG60 from slipping off. The
dimensions are described in the chapter "FHG60 mounting device". → 86
L
CAUTION
Note the following when mounting the device
‣
The mounting device must be installed in such a way as to withstand the weight of the
Gammapilot M under all anticipated operating conditions.
‣
For measuring lengths of 1
600 mm (63 in) or more, two retainers (vertical) or three
retainers (horizontal) must be used.
‣
The use of the bracket or a support unit fitted by the customer is mandatory if the
device is mounted in a vertical position. Otherwise sufficient stability and support for
the Gammapilot M is not ensured.
‣
For stability reasons the mounting version with the terminal housing positioned at the
top should only be used in exceptional cases.
‣
Clamping solution for pipe mounting required by the customer. Do not use supplied
mounting clamps for pipe. Supplied retainers can be used for FMG60
(see picture E)
.
‣
To prevent damage to the tube of the Gammapilot M, the maximum torque that can be
applied to tighten the retainer screws is 6 Nm (4.42 lbf ft).
9.5.3
Ordering information
Detailed ordering information is available from the following sources: