Gammapilot M FMG60 FOUNDATION Fieldbus
Commissioning
Hauser
59
7.2
Switching on the device
Error messages A165 "Electronics defect" and A635 "Present date not defined"
The Gammapilot M contains 2 real-time clocks for decay compensation, which are
permanently compared with one another for safety reasons. In order to bridge voltage
interruptions, the clocks are buffered with a capacitor. To ensure that the clocks work
correctly and retain the date in the event of a voltage interruption, this capacitor must
have a minimum charge. If the A165 "Electronics defect" or A635 "Present date not
defined" error message appears
after switching on
the Gammapilot M, this may
indicate that the capacitor is not yet sufficiently charged. In this case, the Gammapilot
M must be operated at the operating voltage for at least 20 to 30 minutes in order to
charge the capacitor. After this, the correct date must be entered. If the error message
persists after this, it can be deleted by switching the Gammapilot M off and then on
again.
After the supply voltage has been switched on, the device is first initialized. Due to
internal memory tests, this takes approx. 2 minutes.
Local display
FMG60
V01.03.06 FF
The following information appears for approximately 5 seconds:
• Device type
• Software version
• Type of communication signal
Language 092
✓
English
Français
Español
On the first power-up, select the language for the text on the display.
Select the language with the and keys. Press twice to confirm your selection.
Group selection
✓
Basic setup
Calibration
Safety settings
The measured value screen then appears.
The basic setup can now be performed. Press to switch to the group selection:
Press again to enter the first function of the "Basic setup" function group