Troubleshooting
Gammapilot M FMG60 FOUNDATION Fieldbus
90
Hauser
10
Troubleshooting
10.1 System error messages
10.1.1 Error signal
Errors occurring during commissioning or operation are signaled in the following way:
• Error symbol, error code and error description on the display and operating module.
• Foundation Fieldbus signal, configurable (
"Output on alarm (*20)
function):
– MAX: +99999
– MIN: -99999
– Hold (the last value is held)
10.1.2 Last error
The last error is displayed in the
"Previous error" (*A1)
function in the
"Diagnostics"
(*A)
function group. This can be deleted in the
"Clear last error" (*A2)
function.
10.1.3 Types of error
Type of error Symbol
Symbol
Alarm (A)
Continuous
The output signal adopts a value that can be set using the "Output on alarm" (*10)
function:
• MAX: +99999
• MIN: -99999
• Hold (the last value is held)
An error message is displayed
Warning (W)
Flashing
The device continues to measure. An error message is displayed (alternating with
the measured value)
10.1.4 Error codes
Code
Error description
Remedial action
A102 Checksum error
Contact Hauser Service
W103 Initializing
Wait for the initialization procedure to finish
A106 Downloading
Wait for the download to finish
A110 Checksum error
Contact Hauser Service
A111 Electronics defect
Switch device off/on;
If the error persists:
contact Hauser Service or replace the transmitter
A113 Electronics defect
Switch device off/on;
If the error persists:
contact Hauser Service or replace the transmitter
A114 Electronics defect
Switch device off/on;
If the error persists:
contact Hauser Service or replace the transmitter
A116 Download error
Repeat download