Gammapilot M FMG60 FOUNDATION Fieldbus
Commissioning
Hauser
61
A
B
C
D
1
A0018107
A
Stand alone.
Only one Gammapilot M is required for measuring ranges up to 2 m (6.6 ft)
For larger measuring ranges, any number of Gammapilot M devices can be interconnected (cascade mode).
The devices are defined by software settings as:
B
Master
C
Slave (s) or
D
End slave
1
4 to 20 mA HART; PROFIBUS PA; FOUNDATION Fieldbus
Options/display:
•
Stand alone:
This option is selected if the Gammapilot M is used as a single, standalone
device.
•
Master:
This option is selected if the Gammapilot M is located at the beginning of a
cascade chain. It then receives the pulses from a connected slave, adds its own pulses
and calculates the measured value from this total.
•
Slave:
This option is selected if the Gammapilot M is located in the middle of a cascade
chain. It then receives the pulses from another connected slave or end slave, adds its
own pulses and transmits this total to the next device (master or slave). The calibration
is finished if this option is selected. If several transmitters are cascaded, the rest of the
calibration procedure is performed on the master only.
•
End slave:
This option is selected if the Gammapilot is located at the end of a cascade
chain. It does not receive pulses from another device but transmits its own pulses to the
next device (master or slave). The calibration is finished if this option is selected. If
several transmitters are cascaded, the rest of the calibration procedure is performed on
the master only.
•
Not defined:
Is displayed if the operating mode has not yet been selected. An option
must be selected in order to continue with the basic setup.
If a "slave" or an "end slave" is connected to "FieldCare", the pulse rate of this device is
displayed in the header instead of the measured value.
7.3.5
"Measurement mode" (*05)
Local display
Measurement mode *05
✓
Level
Switch
Density
This function is used to select the desired measurement mode.