Gammapilot M FMG60 FOUNDATION Fieldbus
Commissioning
Hauser
57
Step
Function
Action/comments
5
Background pulse rate (*12)
Confirm value with
Continue with full and empty calibration/covered and free calibration
"Full" or "Covered" calibration
For details, see "Full and empty calibration or covered and free calibration"
Step
Function
Action/comments
1
• Switch on radiation
• Vessel remains filled
2
Calibration point (*13)
Select "Full/covered" with or ; confirm with
The following function is for level measurement only
3
Value full (*14)
Enter the current level (60% - 100%)
The following functions are for level and point level measurements
4
Calibration (*15)
Select "Start" with or ; confirm with
5
Avg. pulse rate (*11)
Wait until the value is stable, then confirm with
6
Calibration (*15)
Select "Stop/edit" with or ; confirm with
7
Full calibration (*16)
Confirm value with
8
Next point (*19)
Select "Yes" with or ; confirm with
Continue with "Empty" or "Free" calibration
"Empty" or "Free" calibration
For details, see "Full and empty calibration or covered and free calibration"
Step
Function
Action/comments
1
• The radiation remains switched on
• Empty the vessel
Level: max. 40%
Switch (level limit): radiation path free
2
Calibration point (*13)
Select "Empty/free" with or ; confirm with
The following function is for level measurement only
3
Value empty (*17)
Enter the current level (0% - 40%)
The following functions are for level and point level measurements
4
Calibration (*15)
Select "Start" with or ; confirm with
5
Avg. pulse rate (*11)
Wait until the value is stable, then confirm with
6
Calibration (*15)
Select "Stop/edit" with or ; confirm with
7
Empty calibration (*18)
Confirm value with
8
Next point (*19)
Select "No" with or ; confirm with
This completes the calibration.
7.1.3
Calibration for density and concentration measurements
The general principles for calibrating for density and concentration measurements are
described in "General principles"