Wiring
Gammapilot M FMG60 FOUNDATION Fieldbus
26
Hauser
5
Wiring
5.1
Connection compartments
The Gammapilot M has two connection compartments
1
2
A0018082
1
Connection compartment 1
2
Connection compartment 2
Connection compartment 1
Connection compartment for:
• Power supply
• Signal output (depending on the device version)
Connection compartment 2
Connection compartment for:
• Signal output (depending on the device version)
• Pt-100 input (4-wire)
• Pulse input for cascade mode
• Pulse output for cascade mode
• Display and operating module FHX40 (or VU331)
Depending on the device version, the signal output is located in connection
compartment 1 or 2.
Maximum cable length:
• For cascade, 20 m (66 ft) per device
• For Pt-100 2 m (6.6 ft). Temperature measurement should be performed as close as
possible to density measurement.
5.2
Cable entries
The number and type of cable entries depend on the device version ordered. The following
are possible:
• Coupling M20x1.5. sealing range 7 to 10.5 mm (0.28 to 0.41 in)
• Cable entry M20x1.5
• Cable entry G1/2