Commissioning
Gammapilot M FMG60 FOUNDATION Fieldbus
68
Hauser
"Background calibration" (*10)
Local display
Backgr. cal. *10
Stop/edit
Start
This function is used to start the background calibration
Options:
•
Stop/edit
This option must be selected if:
– No background calibration is to be performed but the pulse rate of an existing
background calibration is to be displayed instead.
– The background calibration is to be performed manually.
Once this option has been selected, the Gammapilot M goes to the
"Bgr. pulse rate"
(*12)
function where the existing pulse rate is displayed and can be changed if
required.
•
Start
This option is used to start automatic background calibration. The Gammapilot M
changes to the
"Avg. pulse rate" (*11)
function
"Avg. pulse rate" (*11)
Local display
Avg. pulse rate *11
186 cps
The average (integrated) pulse rate is displayed in this function (after the selection of
"Start" in the previous function). Initially, this value fluctuates (because of the decay
statistics), but reaches an average value over time due to integration. The longer the
averaging is performed, the lower the fluctuations.
c/s
t
"E"
A0018118
When the value is sufficiently stable users can quit the function by pressing . After this,
the Gammapilot M changes to the
"Background calibration" (*10)
function. Here the