±0.6% of reading in recommended range
+2 to -10% of reading outside high recommended range
±5% of reading outside low recommended range
±4% of reading in recommended range
Measures conductivity in the range 0 to 600,000
µ
S/cm
(600mS/cm). Measurement choices are conductivity,
resistivity, total dissolved solids, salinity, and % concen-
tration. The % concentration selection includes the
choice of five common solutions (0-12% NaOH, 0-15%
HCl, 0-20% NaCl, and 0-25% or 96-99.7% H
2
SO
4
).
The conductivity concentration algorithms for these
solutions are fully temperature compensated. Three
temperature compensation options are available:
manual slope (X%/°C), high purity water (dilute sodium
chloride), and cation conductivity (dilute hydrochloric
acid). Temperature compensation can be disabled,
allowing the analyzer to display raw conductivity. For
more information concerning the use and operation of
the contacting conductivity sensors, refer to the product
data sheets.
Note: When two contacting conductivity sensors are
used, Model 1056 can derive an inferred pH value
called pHCalc. pHCalc is calculated pH, not directly
measured pH.(Model 1056-0X-20-30-AN required)
Note: Selected 4-electrode, high-range contacting
conductivity sensors are compatible with Model 1056.
Input filter:
time constant 1 - 999 sec, default 2 sec.
Response time:
3 seconds to 100% of final reading
Salinity:
uses Practical Salinity Scale
Total Dissolved Solids:
Calculated by multiplying
conductivity at 25 ºC by 0.65
Recommended Sensors For Conductivity
All Rosemount ENDURANCE Model 400 series conduc-
tivity sensors (Pt 1000 RTD) and
Model 410 sensor.
1.3 Contacting Conductivity (Code -20 and -30)
Temperature range
0-150ºC
Temperature Accuracy,
Pt-1000, 0-50 ºC
± 0.1ºC
Temperature Accuracy,
Pt-1000, Temp. > 50 ºC
± 0.5ºC
Performance Specifications
Recommended Range – Contacting Conductivity
Temperature Specifications:
ENDURANCE
TM
series of
conductivity sensors
family
4-electrode sensors
Cell Constant Linearity
Cell 0.01
µ
S/cm 0.1
µ
S/cm 1.0
µ
S/cm 10
µ
S/cm 100
µ
S/cm 1000
µ
S/cm 10mS/cm 100mS/cm 1000mS/cm
Constant
0.01
0.1
1.0
4-electrode
0.01
µ
S/cm to 200
µ
S/cm
0.1
µ
S/cm to 2000
µ
S/cm
1
µ
S/cm to 20mS/cm
2
µ
S/cm to 300mS/cm
200
µ
S/cm to 6000
µ
S/cm
2000
µ
S/cm to 60mS/cm
20mS/cm to 600mS/cm
4
Instruction Manual
LIQ-MAN-1056
Section 1.0: Description and Specification
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...