For use with any standard pH or ORP sensor.
Measurement choices are pH, ORP, Redox, ammonia,
fluoride or custom ISE. The automatic buffer recognition
feature uses stored buffer values and their temperature
curves for the most common buffer standards available
worldwide. The analyzer will recognize the value of the
buffer being measured and perform a self stabilization
check on the sensor before completing the calibration.
Manual or automatic temperature compensation is
menu selectable. Change in pH due to process temper-
ature can be compensated using a programmable tem-
perature coefficient. For more information concerning
the use and operation of the pH or ORP sensors, refer
to the product data sheets.
Model 1056 can also derive an inferred pH value called
pHCalc (calculated pH). pHCalc can be derived and
displayed when two contacting conductivity sensors are
used. (Model 1056-0X-20-30-AN)
Performance Specifications (pH Input)
Measurement Range [pH]:
0 to 14 pH
Accuracy:
±0.01 pH
Diagnostics:
glass impedance, reference impedance
Temperature coefficient:
±0.002pH/ ºC
Solution temperature correction:
pure water, dilute
base and custom.
Buffer recognition:
NIST, DIN 19266, JIS 8802, BSI,
DIN19267, Ingold, and Merck.
Input filter:
time constant 1 - 999 seconds, default 4
seconds.
Response time:
5 seconds to 100%
Temperature Specifications:
Performance Specifications (ORP Input)
Measurement Range [ORP]
:
-1500 to +1500 mV
Accuracy:
± 1 mV
Temperature coefficient:
±0.12mV / ºC
Input filter:
time constant 1 - 999 seconds, default 4
seconds.
Response time:
5 seconds to 100% of final reading
RECOMMENDED SENSORS FOR pH:
All standard pH sensors.
RECOMMENDED SENSORS FOR ORP:
All standard ORP sensors.
Temperature range
0-150 ºC
Temperature Accuracy, Pt-100, 0-50 ºC ± 0.5 ºC
Temperature Accuracy, Temp. > 50 ºC
± 1 ºC
1.5 pH/ORP/ISE (Code -22 and -32)
General purpose and high performance pH sensors
Models 396PVP, 399VP and 3300HT
Instruction Manual
LIQ-MAN-1056
Section 1.0: Description and Specification
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...