60
6.7.8 Use Pressure
The display screen for selecting atmospheric pressure
source. The default value is displayed in
bold type
.
Refer to the Oxygen Programming flow diagram to com-
plete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Use Pressure?
At Air Cal
mA Input
6.7.4 Partial Press
The display screen for selecting pressure units for
Partial pressure is shown. This selection is needed if
the specified measurement is Partial pressure. The
default value is displayed in
bold type
. Refer to the
Oxygen Programming flow diagram to complete this
function.
6.7.5 Salinity
The display screen for Entering the Salinity (as parts
per thousand) of the process liquid to be measured is
shown. The default value is displayed in
bold type
.
Refer to the Oxygen Programming flow diagram to
complete this function.
Enter Salinity as ‰
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Partial Press
mm Hg
in Hg
atm
kPa
mbar
bar
6.7.7 Pressure Units
The display screen for selecting pressure units for
atmospheric pressure is shown. This selection is needed
for the display of atmospheric pressure measured by
the onboard pressure transducer on the Oxygen
measurement board. The default value is displayed in
bold type
. Refer to the Oxygen Programming flow
diagram to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Pressure Units
mm Hg
in Hg
atm
kPa
mbar
bar
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Salinity
00.0 ‰
6.7.6 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type
. Refer to the Oxygen Programming flow
diagram to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Input filter
05 sec
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...