103
To calibrate Pulse Flow scroll to the desired item and
press ENTER.
The following sub-sections provide you with the initial
display screen that appears for each calibration rou-
tine. Use the
diagram for Pulse Flow calibration
at
the end of Sec. 7 and the live screen prompts to com-
plete calibration.
.
To calibrate Pulse Flow scroll to the desired item and
press ENTER.
The following sub-sections provide you with the initial
display screen that appears for each calibration rou-
tine. Use the
diagram for Pulse Flow calibration
at
the end of Sec. 7 and the live screen prompts to com-
plete calibration.
.
Simply enter the know K factor provided with the flow
sensor specifications. The screen will return to the
Pulse Flow Cal Menu and the updated K factor will
appear.
7.11.3 Calibration — Freq/Velocity and Pipe
This screen appears after selecting
Freq/Velocity &
Pipe calibration.
After completing the entry of the Freq/Velocity ratio, the
following screen will appear:
7.11.2 Calibration — K Factor
This screen appears after selecting
K Factor.
.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
K Factor: 12.34 p/Gal
Freq/Velocity & Pipe
In Process
Totalizer Control
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
K Factor: 12.34 p/Gal
Freq/Velocity & Pipe
In Process
Totalizer Control
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Freq/Velocity
1
2.34 Hz per ft/sec
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Pipe Diameter
1
0.00 in
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
K Factor
1
2.34 p/Gal
The following screen will appear if the entries are suc-
cessful. The screen will return to the Pulse Flow Cal
Menu.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Freq/Velocity&Pipe
Updated K Factor
12.34 p/Gal
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...