6.5 Toroidal Conductivity Measurement Programming
6.5.1 Description
The section describes how to configure the Model 1056 analyzer for conductivity measurements using
inductive/toroidal sensors. The following programming and configuration functions are covered.
Measure
Sec.
Menu function:
default
Description
Toroidal
conductivi
ty
6.5.2
Model:
228
Select sensor type
6.5.3
Measure:
Conductivity
Select Conductivity, Resistivity, TDS, Salinity or % conc
6.5.4
Range:
Auto
Select measurement Auto-range or specific range
6.5.5
Cell K:
3.00000/cm
Enter the cell Constant for the sensor
6.5.6
Temp Comp:
Slope
Select Temp Comp: Slope, Neutral Salt, or Raw
6.5.7
Slope:
2.00%/°C
Enter the linear temperature coefficient
6.5.8
Ref Temp:
25.0°C
Enter the Reference temp
6.5.9
Filter:
2 sec
Override the default input filter, enter 0-999 seconds
6.5.10
Custom
Setup
Enter 2-5 data points in ppm and µS/cm for custom curves
To configure the toroidal conductivity measurement
board:
1. Press
MENU
2. Scroll down to
Program
. Press ENTER.
3. Scroll down to
Measurement
. Press ENTER.
4. Select
Sensor 1
or
Sensor 2
corresponding to
toroidal conductivity. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the
flow diagram for toroidal conductivity programming
at the end of Sec. 6 and the Model 1056 live
screen prompts for each function to complete configuration and programming.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Configure
Model: 228
Measure: Cond
Range: Auto
Cell K: 3.00000/cm
RTD Offset: 0.00ºC
RTD Slope: 0
Temp Comp: Slope
Slope: 2.00%/°C
Ref Temp: 25.0°C
Filter: 2 sec
Custom Setup
6.5.2 Sensor Model
The display screen for selecting the sensor model is
shown. The default value is displayed in
bold type
.
Refer to the toroidal conductivity Programming flow
diagram to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Model
228
225
226
247
Other
TABLE 6-4. Toroidal Conductivity Measurement Programming
A detailed flow diagram for toroidal conductivity programming is provided at the end of Sec. 6 to guide
you through all basic programming and configuration functions.
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...