44
6.2 pH Measurement Programming
6.2.1 Description
This section describes how to configure the Model 1056 analyzer for pH measurements. The following programming and
configuration functions are covered.
Measure
Sec.
Menu function: default setting
Description
pH
6.2.2
Measurement type: pH
Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
6.2.3
Preamp location: Analyzer
Identify preamp location
6.2.4
Solution temperature correction Off
Select Off, ultra-pure, high pH, custom
6.2.5
Temp coefficient (custom)
Enter the temp coefficient
6.2.6
Resolution: 0.01pH
Select 0.01pH or 0.1pH for pH display resolution
6.2.7
Filter: 4 sec
Override the default input filter, enter 0-999 seconds
6.2.8
Reference Z: Low
Select low or high reference impedance
To configure the pH measurement board:
1. Press
MENU
2. Scroll down to
Program
. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select
Sensor 1
or
Sensor 2
corresponding to
pH. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any function, scroll to the
desired item and press ENTER.
6.2.2 Measurement
The display screen for selecting the measurement is
shown. The default value is displayed in
bold type
.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Configure
Measure: pH
Preamp: Analyzer
Sol’n Temp Corr: Off
T Coeff: -0.029pH/°C
Resolution: 0.01pH
Filter: 4 sec
Reference Z: Low
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Measurement
pH
ORP
Redox
Ammonia
Fluoride
Custom ISE
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the
flow diagram for pH programming
at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.
6.2.3 Preamp
The display screen for identifying the Preamp location is
shown. The default value is displayed in
bold type
.
Refer to the pH/ORP Programming flow diagram to
complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Preamp
Analyzer
Sensor/JBox
TABLE 6-1. pH Measurement Programming
A detailed flow diagram for pH programming is provided at the end of Sec. 6 to guide you through
all basic programming and configuration functions.
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...