53
6.6 Chlorine Measurement Programming
With a Chlorine measurement board installed, Model 1056 can measure any of four variants of Chlorine:
• Free Chlorine
• Total Chlorine
• Monochloramine
• pH-independent Free Chlorine
The section describes how to configure the Model 1056 analyzer for Chlorine measurements.
6.6.1 Free Chlorine Measurement Programming
6.6.1.1 Description
This Chlorine sub-section describes how to configure the Model 1056 analyzer for Free Chlorine measurement
using amperometric chlorine sensors. The following programming and configuration functions are covered:
To configure the chlorine measurement board for free
chlorine:
1. Press
MENU
2. Scroll down to
Program
. Press ENTER.
3. Scroll down to
Measurement
. Press ENTER.
4. Select
Sensor 1
or
Sensor 2
corresponding to
chlorine. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
Measure
Sec.
Menu function: default
Description
Free
Chlorine
6.6.1.2
Measure: Free Chlorine
Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
6.6.1.3
Units: ppm
Select units ppm or mg/L
6.6.1.4
Filter: 5sec
Override the default input filter, enter 0-999 seconds
6.6.1.5
Free Cl Correct: Live
Select Live/Continuous pH correction or Manual
6.6.1.6
Manual pH: 7.00 pH
For Manual pH correction, enter the pH value
6.6.1.7
Resolution: 0.001
Select display resolution 0.01 or 0.001
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Configure
Measure: Free Chlorine
Units: ppm
Filter: 5sec
Free Cl Correct: Live
Manual pH: 7.00 pH
Resolution: 0.001
TABLE 6-5. Free Chlorine Measurement Programming
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...