99
To calibrate temperature:
1. Press the
MENU
button
2. Select
Calibrate
. Press ENTER.
3. Select
Sensor 1
or
Sensor 2
corresponding to
the desired measurement. Press ENTER.
4. Select Temperature. Press ENTER.
The adjacent screen will appear.
Measure
Sec.
Calibration function: default value Description
Temperature
7.9.2
Calibrate
Enter a manual reference temperature for temperature
compensation of the process measurement
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibrate
+025.0°C
The following sub-section provides you with the initial
display screen that appears for temperature calibration.
Use the
flow diagram for Temp calibration
at the
end of Sec. 7 to complete calibration.
7.9.2 Calibration
The adjacent screen will appear during Temperature
Cal.
If the sensor Temperature offset is greater than 5 ºC
from the default value, the following screen will
appear:
You may continue by selecting Yes or suspend this
operation by selecting No.
If the Temp Cal is successful, the screen will return to
the Cal Menu.
Note: To select automatic or manual temp compensation
or to program temperature units as °C or °F, refer to
Sec. 5.3 – Programming Temperature in this manual
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Temp Offset > 5°C
Continue?
No
Yes
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibrate
Cal in progress.
Please wait.
TABLE 7- 11 Temperature Calibration Routine
7.9 Calibrating Temperature
7.9.1 Description
Most liquid analytical measurements require temperature compensation (except ORP). The Model 1056 performs
temperature compensation automatically by applying internal temperature correction algorithms. Temperature cor-
rection can also be turned off. If temperature correction is off, the Model 1056 uses the manual temperature entered by
the user in all temperature correction calculations.
THIS SECTION DESCRIBES HOW TO CALIBRATE TEMPERATURE IN THE MODEL 1056 ANALyzER. THE
FOLLOWING CALIBRATION ROUTINE IS COVERED.
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...