61
6.8 Ozone Measurement Programming
6.8.1 Description
This section describes how to configure the 1056 analyzer for ozone measurement using amperometric ozone sen-
sors. The following programming and configuration functions are covered:
Measure
Sec.
Menu function: default Description
Ozone
6.8.2
Units: ppm Select units ppm, mg/L, ppb, µg/L
6.8.3
Filter: 5sec
Override the default input filter, enter 0-999 seconds
6.8.4
Resolution: 0.001 Select 0.01or 0.001 for display resolution
To configure the Ozone measurement board:
1. Press
MENU
2. Scroll down to
Program
. Press ENTER.
3. Scroll down to
Measurement
. Press ENTER.
4. Select
Sensor 1
or
Sensor 2
corresponding to
Ozone. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll to
the desired item and press ENTER.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Configure
Units: ppm
Filter: 5 sec
Resolution: 0.001
6.8.2 Units
The display screen for selecting measurement units is
shown. The default value is displayed in
bold type.
Refer to the Ozone Programming flow diagram to
complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Units
ppm
mg/L
ppb
µg/L
6.8.3 Filter
The display screen for entering the input filter value in
seconds is shown. The default value is displayed in
bold type
. Refer to the Ozone Programming flow
diagram to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Input filter
05 sec
6.8.4 Resolution
The display screen for selecting display resolution as
0.001 or 0.01 is shown. The default value is displayed
in
bold type
. Refer to the Ozone Programming flow
diagram to complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Resolution
0.001
0.01
The following sub-sections show the initial display screen that appears for each configuration function. Use the
flow diagram for ozone programming
at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.
Note: Ozone measurement boards are detected automatically by the analyzer. No measurement selection is necessary.
TABLE 6-10. Ozone Measurement Programming
A detailed flow diagram for ozone programming is provided at the end of Sec. 6 to guide you through all
basic programming and configuration functions.
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...