63
If TSS data (Total Suspended Solids) calculation is
selected, the following screen will be displayed. Refer
to the Turbidity programming flow diagram to complete
this function.
6.9.4 Enter TSS Data
The display screen for entering TSS Data is shown.
The default values are displayed. Refer to the
Turbidity Programming flow diagram to complete this
function
Note: Based on user-entered NTU data, calculating
TSS as a straight line curve could cause TSS to go
below zero. The following screen lets users know that
TSS will become zero below a certain NTU value.
The following illustration shows the potential for calculated TSS to go below zero
Normal case: TSS is always a positive number when Turbidity is a positive number.
Abnormal case: TSS can be a negative number when Turbidity is a positive number.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Units
ppm
mg/L
none
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
TSS Data
Calculation Complete
Calculated TSS = 0 below
xxxx NTU
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
TSS Data
Pt1 TSS: 0.000ppm
Pt1 Turbid: 0.000NTU
Pt2 TSS: 100.0ppm
Pt2 Turbid: 100.0NTU
Calculate
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...