6.7.3 Units
The display screen for selecting units as ppm , mg/L,
ppb, µg/L, % Saturation, %Oxygen in Gas, or ppm
Oxygen in Gas is shown. The default value is displayed
in
bold type
. Refer to the Oxygen Programming flow
diagram to complete this function.
6.7.2 Oxygen Measurement application
The display screen for programming the measurement
is shown. The default value is displayed in
bold type.
Refer to the Oxygen Programming flow diagram to
complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Type
Water/Waste
Trace Oxygen
BioRx-Rosemount
BioRx-Other
Brewing
Oxygen In Gas
6.7 Oxygen Measurement Programming
6.7.1 Description
This section describes how to configure the Model 1056 analyzer for dissolved and gaseous oxygen measurement
using amperometric oxygen sensors. The following programming and configuration functions are covered:
To configure the Oxygen measurement board:
1. Press MENU
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to Oxygen.
Press ENTER.
The adjacent screen format will appear (factory defaults are
shown). To program any displayed function, scroll to the
desired item and press ENTER.
Measure
Sec.
Menu function: default Description
Oxygen
6.7.2
Type: Water/Waste Select Water/Waste, Trace. BioRx, BioRx-Other, Brew, %O2 In Gas
6.7.3
Units: ppm Select ppm, mg/L, ppb, µg/L, % Sat, %O2-Gas, ppm Oxygen-Gas
6.7.4
Partial Press: mmHg Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial pressure
6.7.5
Salinity: 00.0‰ Enter Salinity as ‰
6.7.6
Filter: 5sec Override the default input filter, enter 0-999 seconds
6.7.7
Pressure Units: bar Select pressure units: mm Hg, in Hg,. Atm, kPa, mbar, bar
6.7.8
Use Press: At Air Cal Select atmospheric pressure source – internal or mA Input
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Configure
Type: Water/Waste
Units: ppm
Partial Press: mmHg
Salinity: 00.0‰
Filter: 5sec
Pressure Units: bar
Use Press: At Air Cal
Custom Setup
The following sub-sections show the initial display screen that appears for each configuration function. Use the
flow diagram for oxygen programming
at the end of Sec. 6 and the Model 1056 live screen prompts for each
function to complete configuration and programming.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Units
ppm
mg/L
ppb
µg/L
% Saturation
Partial Pressure
% Oxygen In Gas
ppm Oxygen In Gas
TABLE 6-9. Oxygen Measurement Programming
A detailed flow diagram for oxygen programming is provided at the end of Sec. 6 to guide you through all basic pro-
gramming and configuration functions.
59
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...