55
6.6.2.2 Measure
The display screen for selecting the measurement is
shown. The default value is displayed in
bold type
.
Refer to the chlorine Programming flow diagram to
complete this function.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Measurement
Free Chlorine
pH Independ. Free Cl
Total Chlorine
Monochloramine
6.6.2.3 Units
The display screen for selecting units as ppm or mg/L
is shown. The default value is displayed in
bold type
.
Refer to the Chlorine Programming flow diagram to
complete this function
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Units
ppm
mg/L
6.6.2 Total Chlorine Measurement Programming
6.6.2.1 Description
This Chlorine sub-section describes how to configure the Model 1056 analyzer for Total Chlorine measurement using
amperometric chlorine sensors. The following programming and configuration functions are covered:
Measure
Sec.
Menu function:
default
Description
Total
Chlorine
6.6.2.2
Measure: Free Chlorine Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
6.6.2.3
Units: ppm
Select units ppm or mg/L
6.6.2.4
Filter: 5sec
Override the default input filter, enter 0-999 seconds
6.6.2.5
Resolution: 0.001
Select 0.01 or 0.001 display resolution
To configure the chlorine measurement board for total
chlorine:
1. Press
MENU
2. Scroll down to
Program
. Press ENTER.
3. Scroll down to
Measuremen
t. Press ENTER.
4. Select
Sensor 1
or
Sensor 2
corresponding to
chlorine. Press ENTER.
The adjacent screen format will appear (factory defaults
are shown). To program any displayed function, scroll
to the desired item and press ENTER.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Configure
Measure: Free Chlorine
Units: ppm
Filter: 5sec
Resolution: 0.001
The following sub-sections provide you with the initial display screen that appears for each configuration function.
Use the flow diagram for chlorine programming at the end of Sec. 6 and the Model 1056 live screen prompts for
each function to complete configuration and programming.
TABLE 6-6. Total Chlorine Measurement Programming
A detailed flow diagram for programming of all chlorine measurements is provided at the end of Sec. 6 to
guide you through all basic programming and configuration functions.
Instruction Manual
LIQ-MAN-1056
Section 6.0: Programming the Measurements
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...