The following sub-sections show the initial display
screen that appears for each calibration routine. Use
the
flow diagram for Conductivity calibration
at the
end of Sec. 7 and the live screen prompts to complete
calibration.
The adjacent screen appears after selecting
Conductivity calibration:
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
Zero Cal
In Process Cal
Cell K: 1.00000/cm
To calibrate toroidal conductivity:
1. Press the
MENU
button
2. Select
Calibrate
. Press ENTER.
3. Select
Sensor 1
or
Sensor 2
corresponding to
Toroidal Conductivity. Press ENTER.
4. Select
Conductivity
. Press ENTER.
The adjacent screen will appear. To calibrate Toroidal
Conductivity or Temperature, scroll to the desired item
and press ENTER
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibrate?
Conductivity
Temperature
Measure
Sec.
Calibration function: default value Description
Toroidal
Conductivity
7.5.2
Cell K: 3.00000/cm Enter the cell Constant for the sensor
7.5.3
Zero Cal
Zeroing the analyzer with the sensor attached
7.5.4
In Process Cal
Standardizing the sensor to a known conductivity
A detailed flow diagram is provided at the end of Sec. 7 to guide you through the calibration routines.
TABLE 7-4
Toroidal Conductivity Calibration
7.5 Toroidal Conductivity Calibration
7.5.1 Description
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instru-
ment, or standardizing to some known reference (such as a conductivity standard). This section contains procedures
for the first time use and for routine calibration of the Model 1056 analyzer.
THIS SECTION DESCRIBES HOW TO CALIBRATE THE MODEL 1056 WITH AN ATTACHED INDUC-
TIVE/TOROIDAL CONDUCTIVITy SENSOR. THE FOLLOWING CALIBRATION ROUTINES ARE COVERED
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...