93
The following sub-sections provide you with the initial
display screen that appears for each calibration
routine. Use the
flow diagram for Chlorine
calibration
at the end of Sec. 7 and the live screen
prompts to complete calibration.
The adjacent screen appears after selecting
pH-inde-
pendent free chlorine calibration:
7.6.4.2 zeroing The Sensor
The adjacent screen will appear during Zero Cal
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Zeroing
Wait
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Calibration
Zero Cal
In Process Cal
The adjacent screen may appear if In Zero Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
The adjacent screen will appear if In Zero Cal is suc-
cessful. The screen will return to the Amperometric
Cal Menu.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Sensor zero failed
Press EXIT
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
Zero Cal
Sensor zero done
If the In Process Cal is successful, the screen will
return to the Cal sub-menu.
The adjacent screen may appear if In Process Cal is
unsuccessful. The screen will return to the
Amperometric Cal Menu.
7.6.4.3 In Process Calibration
The following screen will appear prior to In Process Cal
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Calibration
Error
Press EXIT
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
S
N
InProcess Cal
Wait for stable
reading.
Instruction Manual
LIQ-MAN-1056
Section 7.0: Calibration
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...