1.11 Turbidity (Code 27 and 37)
The 1056 instrument is available in single and dual tur-
bidity configurations for the Clarity II
®
turbidimeter. It is
intended for the determination of turbidity in filtered
drinking water. The other components of the Clarity II
turbidimeter – sensor(s), debubbler/measuring cham-
ber(s), and cable for each sensor must be ordered
separately or as a complete system with the Model
1056.
The 1056 turbidity instrument accepts inputs from both
USEPA 180.1 and ISO 7027-compliant sensors
When ordering the Model 1056 turbidity instrument, the
02 (24VDC power supply) or the 03 (switching
115/230VAC power supply) are required. Both of these
power supplies include four fully programmable relays
with timers.
Note:
Model 1056 Turbidity must be used with Clarity
II sensor, sensor cable and debubbler.
Performance Specifications
Units:
Turbidity (NTU, FTU, or FNU); total suspended
solids (mg/L, ppm, or no units)
Display resolution-turbidity:
4 digits; decimal point
moves from x.xxx to xxx.
Display resolution-TSS:
4 digits; decimal point moves
from x.xxx to xxxx
Calibration methods:
user-prepared standard, com-
mercially prepared standard, or grab sample. For total
suspended solids user must provide a linear calibration
equation.
Inputs:
Choice of single or dual input, EPA 180.1 or
ISO 7027 sensors.
Field wiring terminals:
removable terminal blocks for
sensor connection.
Accuracy after calibration at 20.0 NTU:
0-1 NTU ±2% of reading or 0.015 NTU, whichever is
greater.
0-20 NTU: ±2% of reading.
Clarity ll Turbidimeter
10
Instruction Manual
LIQ-MAN-1056
Section 1.0: Description and Specification
April 2017
Summary of Contents for Rosemount 1056
Page 4: ...Quick Start Guide Figure A Quick Start Guide ...
Page 5: ...Quick Reference Guide Figure B Model 1056 Menu Tree ...
Page 28: ...20 Instruction Manual LIQ MAN 1056 Section 2 0 Installation April 2017 ...
Page 50: ...42 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 84: ...76 Instruction Manual LIQ MAN 1056 Section 6 0 Programming the Measurements April 2017 ...
Page 122: ...114 Instruction Manual LIQ MAN 1056 Section 7 0 Calibration April 2017 ...
Page 124: ...116 Instruction Manual LIQ MAN 1056 Section 8 0 Return of Material April 2017 ...